Installation of 300 MHz Bruker DPX System

After energizing all the coils a shunt is installed to temporarily complete the circuit. Then the power supply is disconnected from the magnet. The charging rod is removed and the shunt plug is installed. If everything goes smoothly the coils remain energized.
The final steps involve shimming with the room temperature shims to achieve the published instrument specifications on line shape, resolution, and signal-to-noise.